CSL Part Number: 104032
Product Overview: The Thermotron 8800 Temperature Chamber Controller is a single-board computer controller used in environmental test chambers for temperature cycling and thermal testing applications. It runs Windows Embedded operating systems and uses industrial motherboards with Intel Celeron processors. The unit features VGA display output, USB ports, serial communications (COM ports), and connects to temperature chambers via multiple I/O interfaces. It includes solid-state storage (SSD or NVMe), a CMOS battery for BIOS retention, and requires 15 VDC power input. The controller manages chamber temperature profiles, monitors sensors and transducers, and provides touchscreen compatibility in some configurations.
Category: Temperature Control
Manufacturer: Thermotron Industries
Manufacturer Synonyms: Thermotron
Manufacturer Part Number: 8800

Common Failures

Several units experienced power and boot issues, often linked to components such as CMOS batteries, motherboards, and storage drives. Symptoms included units not powering on, no display output, and instability in operation. Defective components identified included motherboards, hard drives, RAM, and USB connectors.


If the unit is running Windows 10 IOT (identified by the Microsoft sticker on the unit) and the hard drive has failed, then Thermaltron must be contacted directly. This is because the software licenses are stored on the hard drive for this version of the control. Some symptoms of this type of failure are: "blue screen of death", disappearing data, and the message "no bootable device" at the time of startup.

Repair Approach

Repairs typically involved replacing defective parts such as CMOS batteries and motherboards. Adjustments were made to BIOS settings to support devices booting when power is applied. For boot failures resulting from memory or hard drive issues, components were replaced or repaired to restore functionality.

Testing & Quality Assurance

Each repaired device underwent thorough testing to verify operational status. This included testing procedures, burn-in tests, and verification that communication interfaces were functional. For specific devices, BIOS configurations were adjusted to support power-on boot functionality.

Service and repair your broken Thermotron Industries part


We can repair your Thermotron Industries model 8800 in our facility. Let our skilled technicians get your Thermotron Industries back in operation quickly.


Step 1 - Click the green 'Repair Now' button above to add the part and then complete the RMA.

Step 2 - The RMA form will be sent to you via email. Print the RMA form and put it in a box with the item(s) for repair.

Step 3 - Ship or mail the box to the address listed on the RMA.